Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected systems. Affected PCs: FlexVision PC Allura Xper Model Name - Model Number: Allura Xper FD10 722003 722010 722026; Allura Xper FD10/10 722005 722011 722027; Allura Xper FD20 722006 722012 722028; Allura Xper FD20 Biplane 722008 722013; Allura Xper FD20 Biplane OR Table 722025; Allura Xper FD20 OR Table 722023 722035; Allura Xper FD20/10 722029; Allura Xper FD20/15 722058; Allura Xper FD20/15 OR Table 722059; Allura Xper FD20/20 722038; Allura Xper FD20/20 OR Table 722039;

FDA Recall Z-1144-2024 · Posted 2024-02-16

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2024-01-02
Product Quantity1,993 US; 2,662 OUS

Reason for Recall

Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.

Firm Action

On January 2, 2024, the firm notified affected consignees via URGENT Medical Device Correction letters. These letters covered issues with three components of the PCs that may not perform as intended and which may therefore lead to delay in treatment: 1) DIMMs, 2) Disk Bay, and 3) Framegrabber Card. This recall record is for the Framegrabber Card issue. Customers should circulate the notice to all users and establish an emergency protocol for if you experience the issue during a procedure. Customers were provided with monitoring activities to undertake while waiting for their system to be corrected. Philips will be replacing the three components in all affected systems. Philips will contact customers to schedule a visit to replace the affected components. If you need additional information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center).

Root Cause

Process design

Distribution Pattern

Worldwide - US Nationwide distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑