Terminated

Isoloc software Versions 6.5 and 4.5, Part Number NT-NW-425-520. Software for image guided radiation therapy.

FDA Recall Z-1145-2007 · Posted 2007-08-08

Recall StatusTerminated
Product CodeIYE
Date Initiated2006-01-02
Date Terminated2011-04-19
Product Quantity132

Reason for Recall

Isoloc version 6.5 and 4.5 software issued a false warning message regarding Gantry angles when set to "Anatomic Landmark" setting.

Firm Action

Software to fix the described problem was written and customers contacted by e-mail on 01/06/2006 that a software patch, titled 01022006 was available to their customers on the firm's internet website.

Root Cause

Software design

Distribution Pattern

Nationwide, Korea, Spain, and UK.