ISOLOC 6.5 Software Part Number NT-NW-425-520. Software for radiotherapy treatment.

FDA Recall Z-1146-2007 · Posted 2007-08-08

Recall StatusTerminated
Product CodeIYE
Date Initiated2006-04-07
Date Terminated2011-04-19
Product Quantity141 copies

Reason for Recall

This correction was for several software bugs which caused image format problems; incorrect ruler display; software crashes; failure to update pixel spacing; display of unattainable treatment table positions; failure to update gantry angles; wrong shifts for two of three axes of treatment table movement.

Firm Action

Software to fix the problems was written and software users were notified of the availability by e-mail on 04/07/2006. They could download the fixed software from the firm's internet website.

Root Cause

Software design

Distribution Pattern

Customers in USA, Korea, Spain and UK.