ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software.
FDA Recall Z-1147-2007 · Posted 2007-08-08
Advertisement
Recalling FirmNorthwest Medical Physics Equipment Inc
Recall StatusTerminated
Product CodeIYE
Date Initiated2005-05-31
Date Terminated2011-04-19
Product Quantity85
Reason for Recall
When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display values ten times greater than the correct values.
Firm Action
Software to fix the described problems was written and customers contacted by e-mail on 05/31/2005, that a software patch, titled 05312005 was available on the firm's internet website.
Root Cause
Software design
Distribution Pattern
Worldwide, including USA, Korea, Spain, and UK.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.