Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

ConMed 50L abdominal insufflator, Catalog Number 72-00258-0 / GS2000 - Product Usage: The device will maintain an expanded cavity (pneumoperitoneum) to allow the insertion of a variety of surgical instruments through the abdominal wall for visual examination of the interior or for surgery (laparoscopic surgery).

FDA Recall Z-1147-2021 · Posted None

Recall StatusTerminated
Product CodeHIF
Date Initiated2021-01-15
Date Terminated2022-10-31
Product Quantity25 units

Reason for Recall

An internal tubing component in the flow path was made with incorrect material which was not tested for use with this device.

Firm Action

The firm initiated the recall on 01/15/2021 by email to its direct consignees. The notice requested the consignee cease distribution or use and return the product to NTI. It also directed the consignee to notify their customers if the product was further distributed.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the states of FL and OH.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑