Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197
FDA Recall Z-1148-2020 · Posted 2020-03-02
Advertisement
Recalling FirmStryker Corporation
Recall StatusTerminated
Product CodeMBI
Date Initiated2019-07-19
Date Terminated2024-02-02
Product Quantity87 devices
Reason for Recall
Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.
Firm Action
On July 23, 2019, a "Urgent Medical Device Recall Notification" was emailed to sales representatives. In addition to informing consignees on the cause of the recall, the following action was requested:
1. Check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your accounts. Response is required by August 15, 2019.
2.If affected product is found in your inventory, segregate the product and call Stryker customer service at 1-866-596-2022 or email [email protected] to arrange for product return and issuance of credit or replacement (upon availability). Also fill out Attachment A and return to [email protected].
3.If affected product is found at account(s), please complete both steps below: a.Provide attached memo (Attachment B) to the account(s) to acknowledge the removal of devices and send completed memo back to [email protected]. b.Segregate the product and call Stryker customer service at 1-866-596-2022 or email [email protected] to arrange for product return and issuance of credit or replacement (upon availability).
4.If no product is found in your inventory or at your account(s), complete acknowledgement form on Attachment A and return it back to [email protected].
Root Cause
Nonconforming Material/Component
Distribution Pattern
US: GA, NJ, NY, KY, MD, TX, UT, AZ, IL, ID, CO, OH, FL, AK, CA, NV, AR, ME, WI
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.