Steris CO2MPACT Endoscopic Insufflator system, Catalog Number 72-00310-0 / 710300 - Product Usage: designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope.

FDA Recall Z-1148-2021 · Posted None

Recall StatusTerminated
Product CodeFCX
Date Initiated2021-01-15
Date Terminated2022-10-31
Product Quantity45 units

Reason for Recall

An internal tubing component in the flow path was made with incorrect material which was not tested for use with this device.

Firm Action

The firm initiated the recall on 01/15/2021 by email to its direct consignees. The notice requested the consignee cease distribution or use and return the product to NTI. It also directed the consignee to notify their customers if the product was further distributed.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the states of FL and OH.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.