InterFuse Tail Traction Tool, Part Number 9116. The InterFuse Tail Traction Tool is an optional Class I non-sterile reusable instrument that can be used to assist with module locking of the InterFuse S or InterFuse T implant system. The system has a modular design to allow the device to be assembled within the disc space to minimize the surgical window, while providing a footprint that approximates the excised nucleus.

FDA Recall Z-1151-2017 · Posted None

Recall StatusTerminated
Product CodeLXH
Date Initiated2017-01-13
Date Terminated2017-10-05
Product Quantity63 pieces

Reason for Recall

Fit issues have been identified with the Tail Traction Tool (PN 9116) when used with the Interfuse T tailed implant. The InterFuse T Tail may not fit into the Thru Hole in the instrument.

Firm Action

Consignees were sent via email a VTI "Advisory Notice" letter. The letter listed the Concern, Impact, Alternate Methods, Affected Lots, and Your Responsibility. Consignees were requested to acknowledge receipt of this notice, Discontinue use of the Tail Traction Tool when implanting the InterFuse T tailed implant, Notify individuals that may potentially use the device and alternate module assembly methods. Advise consignees to contact Customer Service (1-877-912-5401 in US , +-952-912-5400 outside US or [email protected] to obtain a replacement instrument.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US: IL, WI, OH, NY, MT, FL, MD, IN, TN, TX, MN. OUS: Italy, Taiwan Spain, Belgium, Canada, Netherlands, Sweden, Switzerland, Portugal, Brazil. Germany, Ireland,