ISOLOC Software Versions 6.0 and 4.2. Part Number NT-NW-425-520. Image guided radiotherapy treatment software.
FDA Recall Z-1152-2007 · Posted 2007-08-08
Advertisement
Recalling FirmNorthwest Medical Physics Equipment Inc
Recall StatusTerminated
Product CodeIYE
Date Initiated2005-04-28
Date Terminated2011-04-20
Product Quantity84
Reason for Recall
When ISOLOC version 6.0 and 4.2 distance unit is set to centimeter, the localization coordinates were rounded to the nearest millimeter which could affect positioning by 0.5 mm.
Firm Action
Software to fix the described problems was written and customers contacted by e-mail on 04/28/2005, that a software patch, titled 04282005, was available on the firm's internet website.
Root Cause
Software design
Distribution Pattern
Worldwide, including USA, Korea, Spain, and UK.