ARROWg+ard Blue PLUS catheter, Catalog Number: ASK-45854-NG Product Usage: The ARROWg+ard Blue PLUS catheter permits venous access to central circulation by way of subclavian, jugular, and femoral veins. The ARROWg+ard technology is intended to provide protection against catheter-related bloodstream infections (CRBSIs). It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use (> 30 days). Clinical effectiveness of the ARROWg+ard Blue PLUS catheter in preventing CRBSIs compared to the original ARROWg+ard Blue catheter has not been studied.
FDA Recall Z-1152-2018 · Posted 2018-03-21
Advertisement
Recalling FirmArrow International Inc
Recall StatusTerminated
Product CodeFOZ
Date Initiated2017-12-21
Date Terminated2020-04-20
Product Quantity420
Reason for Recall
Certain lots may be missing information at the bottom of the lid including the lot number, expiration date, and Unique Device Identification (UDI) information.
Firm Action
Arrow International sent Notification letters to DISTRIBUTORS that include the following instructions: 1. Discontinue distribution and quarantine any affected products. 2. Inspect affected products to identify whether they have the lot number, expiration date, and UDI information. Products missing this information should be returned. 3. Using the provided Customer Letter and Acknowledgement Form templates, communicate this recall to any customers who have received affected products. Customers should return mislabeled products. Products that are properly labeled should not be returned. 4. Any customers with mislabeled products should return them with a completed Acknowledgement Form. 6. If you and customers have no affected stock please complete and return the Acknowledgment Form.
Notification letters to RISK MANAGERS include the following instructions: 1. Discontinue use and quarantine any affected products. 2. Inspect affected products and identify whether they have the lot number, expiration date, and UDI information. Products missing this information should be returned to the Distributor. 3. Complete the Acknowledgement Form whether or not there is affected product that must be returned.
For questions contact your local sales representative or Customer Service at 1-866-246-6990.
Root Cause
Process control
Distribution Pattern
United States in the state of FL.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement