ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position patients) for high accuracy radiotherapy.

FDA Recall Z-1153-2007 · Posted 2007-08-08

Recall StatusTerminated
Product CodeIYE
Date Initiated2005-04-11
Date Terminated2011-04-19
Product Quantity84

Reason for Recall

When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm.

Firm Action

Software to fix the described problem was written and customers contacted by e-mail on 04/11/2005, that a software patch, titled 04112005, was available to their customers on the firm's internet website.

Root Cause

Software design

Distribution Pattern

Worldwide, including USA, Korea, Spain, and UK.