ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position patients) for high accuracy radiotherapy.
FDA Recall Z-1153-2007 · Posted 2007-08-08
Advertisement
Recalling FirmNorthwest Medical Physics Equipment Inc
Recall StatusTerminated
Product CodeIYE
Date Initiated2005-04-11
Date Terminated2011-04-19
Product Quantity84
Reason for Recall
When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm.
Firm Action
Software to fix the described problem was written and customers contacted by e-mail on 04/11/2005, that a software patch, titled 04112005, was available to their customers on the firm's internet website.
Root Cause
Software design
Distribution Pattern
Worldwide, including USA, Korea, Spain, and UK.