Velocity Disposable Bipolar Ablators - Product Usage: intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures.
FDA Recall Z-1153-2021 · Posted None
Advertisement
Recalling FirmBovie Medical Corporation
Recall StatusOpen, Classified
Product CodeGEI
Date Initiated2021-02-01
Product Quantity5894 units
Reason for Recall
There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration.
Firm Action
The firm sent "URGENT MEDICAL DEVICE RECALL" (FA-2021-1) and response form to customers on February 1, 2021.
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in the states of CA and FL.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K161558Bovie 3.3mm Disposable Bipolar Ablator 90°, Aspirating and Non-Aspirating, Bovie 2.4mm Disposable Bipolar Ablator 55°, Non-Aspirating, Bovie 1.8mm Disposable Bipolar Ablator 60°, Non-AspiratingBovie Medical Corporation · 2016-07-29K152777Bovie Disposable Bipolar AblatorBovie Medical Corporation · 2016-01-15
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.