Velocity Disposable Bipolar Ablators - Product Usage: intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures.

FDA Recall Z-1153-2021 · Posted None

Recall StatusOpen, Classified
Product CodeGEI
Date Initiated2021-02-01
Product Quantity5894 units

Reason for Recall

There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration.

Firm Action

The firm sent "URGENT MEDICAL DEVICE RECALL" (FA-2021-1) and response form to customers on February 1, 2021.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of CA and FL.