Terminated

Power Processor Aliquot Tube Labels Part 800255

FDA Recall Z-1154-04 · Posted 2004-07-23

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Date Initiated2003-06-23
Date Terminated2004-08-30
Product Quantity170 rolls

Reason for Recall

Labels are missing an expiration date and storage instructions, increasing the likelihood that they will not be used by the expiration date or may be stored under conditions that would cause adhesive failure and loss of identity of sample or aliquot during testing.

Root Cause

Other