Stericover/ Sterile Disposible Polyethylene Drape ---Universal Video Camera/ Laser Arm Equipment Cover; 7'' x 96''.
FDA Recall Z-1154-06 · Posted 2006-06-17
Advertisement
Recalling FirmAbbott Spine
Recall StatusTerminated
Product CodeKKX
Date Initiated2006-05-15
Date Terminated2007-09-25
Product Quantity326 units.
Reason for Recall
Non-sterility-Distributor testing of outside manufacturer's product (Sterimed, Inc) revealed positive sterility test results.
Firm Action
Firm notified sales agents of recall via telephone on 05/15/06 and requested product returns for destruction
Root Cause
Other
Distribution Pattern
NATIONWIDE-Drapes consigned to sales agents located in CA, FL, IL, MD, MI, OK, OH, TX, and WI.