ISOLOC Software Version 4.2, Part Number NT-NW-425-520. Image guided radiation treatment software.

FDA Recall Z-1155-2007 · Posted 2007-08-09

Recall StatusTerminated
Product CodeIYE
Date Initiated2004-11-30
Date Terminated2011-04-19
Product Quantity83

Reason for Recall

When ISOLOC Version 4.2 unit is set to centimeter and the user clicks the Reenter button in the digitizer window, the coordinates of the markers and target in the anatomic imaging system erroneously displays values ten times greater than the correct values.

Firm Action

Software to fix the described problem was written and customers contacted by e-mail on 11/30/2004, that a software patch, titled 11302004, was available to their customers on the firm's internet website.

Root Cause

Software design

Distribution Pattern

Worldwide, including USA, Korea, Spain, and UK.