GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System
FDA Recall Z-1155-2009 · Posted 2009-09-24
Advertisement
Recalling FirmGe Healthcare
Recall StatusTerminated
Product CodeJAA
Date Initiated2009-02-19
Date Terminated2011-07-22
Product Quantity554 units
Reason for Recall
The alarm timer is reset after each exposure and therefore never reaches the five minute alarm state unless one exposure exceeds 5 minutes in duration.
Firm Action
GE Healthcare sent a customer notification out to all sites that have potentially affected systems. GEHC Surgery will also install a software correction into the affected systems in order to resolve this issue at no charge to the customer. Until the GEHC OEC Fluorostar 7900 has received the corrective software installed, they are encouraging facilities to exercise caution in monitoring the amount of cumulative fluoroscopy time when the affected system is in use.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Worldwide Distribution