Piccolo Comprehensive Metabolic Panel Reagent Disc, 10-pack box, for use with the Piccolo Blood Chemistry Analyzer or Piccoloxpress Chemistry Analyzer, Catalog No. 400-0028, manufactured by Abaxis Inc., Union City, CA Medical device for use in the in vitro quantitative determination of alanine aminotransferase (ALT), albumin, alkaline Phosphatase (ALP), aspartate aminotransferase, calcium, chloride, creatinine, glucose, potassium, sodium, total bilirubin, total carbon dioxide, total protein, and blood urea nitrogen in heparinized whole blood, heparinized plasma or serum.

FDA Recall Z-1155-2010 · Posted 2010-03-16

Recalling FirmAbaxis Inc
Recall StatusTerminated
Product CodeJFY
Date Initiated2009-11-16
Date Terminated2010-12-14
Product Quantity1010 boxes

Reason for Recall

Wrong barcode applied to a lot of items, which may result in incorrect calibration factors and results.

Firm Action

An "URGENT: PRODUCT RECALL NOTICE" dated November 16, 2009, was sent to customers. The notification described the product, issue/impact, and the action/resolution to be taken by the customers. The customers are to discontinue use of the specified lot, contact Abaxis for immediate replacement and complete and return the enclosed Response form within 5 working days. They may fax the form to Abaxis Technical Support at 1-510-315-3140. The customer are requested to ensure all laboratory staff are informed about this notice and retain this notification as a part of their laboratory Quality System documentation. The firm is currently contacting all direct shipment customers. If you have any questions regarding this Product Recall Notice, please contact Abaxis Medical Technical Support at 800-822-2947, extension 6505, or at [email protected].

Root Cause

Labeling mix-ups

Distribution Pattern

Worldwide distribution: USA and Canada, EU and Asia