NurseMate with Physio Module; Remote Review, Monitoring and Charting Station; USA EPMedSystems, Inc. Cooper Run Executive Park, 575 Route 73 North, Unit D, West Berlin, NJ 08090. Outer box labeled as Smiths Medical. The NurseMate Remote Review Station is an integrated review station for the EP WorkMate Computer recording system that allows a separate user to review and edit patient study date stored on the SP Work Mate System and monitor patient data from the EP WorkMate during a Patient study.
FDA Recall Z-1156-2009 · Posted 2009-04-10
Advertisement
Recalling FirmSt. Jude Medical
Recall StatusTerminated
Product CodeDQK
Date Initiated2009-02-13
Date Terminated2009-05-04
Product Quantity4 units
Reason for Recall
Installation problem: there is a risk that the monitor may not be adequately secured to the stand, resulting in the monitor falling off which may cause injury.
Firm Action
St. Jude Medical received a recall notice dated January 16, 2009 from its supplier, Smiths Medical PM, Inc. St. Jude Medical prepared and sent their own recall letter dated February 13, 2009 to affected customers. The recall letter described the issue and possible associated risks and provided instructions to customers. A St. Jude Medical representative will contact customers to facilitate replacement of the fasteners. Questions should be directed to your St. Jude Medical sales representative.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide Distribution -- CA, NV, and FL.