Medicrea TLIF-B Holder, short intended for Intervertebral body fusion device Catalog: Number A24200010
FDA Recall Z-1156-2020 · Posted None
Advertisement
Recalling FirmMedicrea International
Recall StatusTerminated
Product CodeMAX
Date Initiated2019-10-02
Date Terminated2021-02-08
Product Quantity14 units
Reason for Recall
Inner draw rod, which threads into the cage may potentially be separated from the proximal knob of the inserter
Firm Action
Medicrea Issued Voluntary Product Recall dated 10/3/19 stating reason for recall, health risk and action to take: 1. Immediately check your internal inventory and quarantine all subject devices pending return to MEDICREA USA
2. Circulate this notification internally to all interested/affected parties.
3. Inform MEDICREA USA if any of the subject devices have been distributed to other organizations. 4. Please inform M EDICREA of any adverse events concerning the use of the subject devices. Complete the attached customer response form. Contact person for this action is given below:
Tim Huntington MEDICREA USA
QA Manager
Tel : 646-490-2426
Mail : [email protected]
Root Cause
Process control
Distribution Pattern
OH, VA