Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

BIOPLEX 2200 EBV IgG KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1250 - manufactured by Bio-Rad Laboratories in Redmond, WA 98052. The product is multiflow immunoassay intended for the quantitative detection of IgG antibodies to three (3) seperate EBV antigens in human serum. They are: 1) Epstein-Barr Virus Nuclear Antigen-1 (EBV-NA-1); 2) Viral Capsid Antigen (EBV VCA); and 3) Early Antigen diffuse (EBV EA-D).

FDA Recall Z-1157-2008 · Posted 2008-06-11

Recalling FirmBio-Rad Laboratories
Recall StatusTerminated
Product CodeLSE
Date Initiated2007-11-13
Date Terminated2011-05-23
Product Quantity1,020 kits

Reason for Recall

False negative results due to reagent packs exhibiting low signal.

Firm Action

On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑