Medline Contro-Bulb Syringe, REF DYND20125; bulb irrigation syringe
FDA Recall Z-1157-2024 · Posted 2024-09-30
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeKYZ
Date Initiated2024-01-04
Product Quantity4831 units
Reason for Recall
Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001).
Firm Action
Medline Industries, LP issued an IMMEDIATE ACTION REQUIRED notice to its consignees on 01/04/2024 via USPS First Class. The notice explained the problem with the device, risk involved, and requested the consignee quarantine then destroy the affected product. Upon receipt of the completed destruction form, firm will issue credit, if applicable. Consignee was directed to notify all parties of this action to whom the product was distributed. For questions, contact 866-359-1704
Root Cause
Under Investigation by firm
Distribution Pattern
US, Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.