Terminated

Leksell Gamma Knife Perfexion, Article #715000. Teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.

FDA Recall Z-1158-2010 · Posted 2010-03-17

Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeIWB
Date Initiated2007-07-02
Date Terminated2010-03-25
Product Quantity19

Reason for Recall

Need to modify the closing speed of the shielding doors in the event of an emergency exit

Firm Action

A field action (FCO TN-07-0007) has been performed to modify the shielding door speed on all sites in clinical use. A second Field Change Order 200 069 "Replacement of Door Servo Board" dated 9/16/08 was released.

Root Cause

Other

Distribution Pattern

Worldwide Distribution -- USA, including states of AR, AZ, CA, FL, IL, MI, MN, MS, NJ, OH, OR, PA, UT, VA and countries of France and United Kingdom.