cobas infinity core, Material No. 7154003001. The device is a calculator/data processing module, for clinical use.

FDA Recall Z-1158-2020 · Posted None

Recall StatusTerminated
Product CodeJQP
Date Initiated2020-01-07
Date Terminated2020-10-15
Product Quantity26 accounts

Reason for Recall

Alarm flags that have a "<" symbol in front of them are not displayed on the cobas infinity Validation screen. This issue affects only customers that are doing manual validation on the cobas infinity Validation screen.

Firm Action

On 1/7/2020, the firm distributed Urgent Medical Device Correction letters to affected consignees. Customers were asked to do the following: " Be aware of the incorrect display of the alarm flags on the cobas infinity Validation screen if an alarm has a < symbol in front of it and if doing manual validations. " It is not recommended to use the alarm mapping function on the cobas infinity until your system is patched. If your laboratory does use it, please call the Roche Support Network Customer Support Center at 1-800-526-2272 for further information. " Complete the customer response form and return it to the recalling firm. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272 if you have questions about the information contained in this UMDC.

Root Cause

Under Investigation by firm

Distribution Pattern

Distributed to customers in AK, AL, CA, GA, HI, IA, IL, KY, MA, MI, MN, MO, MS, MT, NJ, NY, OR, SC, and TN.