Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM: a) 100 test - Catalog No. 665-1150; and b) the 5,000 tests - Catalog No. 665-1155 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is used as an aid in the diagnosis of systemic autoimmune diseases.

FDA Recall Z-1159-2008 · Posted 2008-06-11

Recalling FirmBio-Rad Laboratories
Recall StatusTerminated
Product CodeLKJ
Date Initiated2007-11-13
Date Terminated2011-05-23
Product Quantity8,804 kits

Reason for Recall

False negative results due to reagent packs exhibiting low signal.

Firm Action

On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑