Elekta SL Series/Precise Digital Accelerator. The product is used with radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.

FDA Recall Z-1159-2010 · Posted 2010-03-18

Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeIYE
Date Initiated2005-10-01
Date Terminated2010-03-25
Product Quantity315 units

Reason for Recall

Customer reported larger than expected variations in wedge factory with gantry rotation.

Firm Action

Important Notice A280 "Wedge Matching Procedure" (dated 11/18/05) was sent to all effected customers. The Important Notice described the issue, warnings and cautions and corrective instructions/Maintenance to take to eliminate occurrence. Direct questions to your local Elekta Limited representative.

Root Cause

Device Design

Distribution Pattern

United States, Canada and Mexico.