Terminated

Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of insulin levels in human serum and plasma (EDTA) using the Access Immunoassay Systems. Beckman Coulter, Inc., Chaska, MN 55318.

FDA Recall Z-1160-2008 · Posted 2008-06-25

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeCFP
Date Initiated2007-09-19
Date Terminated2012-05-09
Product Quantity8,205 units in the US; 332 in Canada

Reason for Recall

False negative results when used to test serum samples (as compared to plasma samples).

Firm Action

On September 19, 2007, Beckman Coulter mailed to its consignees an Urgent Product Corrective Action Letter informing them that the affected lots of the product were giving false negative results when used to test serum samples (as compared to plasma samples). The letter instructed consignees to discontinue using the affected lots of the product, and to re-test using an EDTA plasma sample if there was any discordance between the results of the serum sample (if it was obtained using an affected lot of the product) as compared with the clinical presentaion of the patient and the results of other diagnostic tests. The letter also instructed consignees to return, within 10 days, the Response Sheet .

Root Cause

Other

Distribution Pattern

Worldwide: including USA and Canada.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.