Terminated

B Braun 9F (O/S) Locking Tearway Introducers non-sterile. Catalog number 614008. The product is packaged as 50 bulk nonsterile units per pouch, 2 pouches per shipping carton.

FDA Recall Z-1163-06 · Posted 2006-06-24

Recall StatusTerminated
Product CodeDYB
Date Initiated2006-05-08
Date Terminated2006-08-22
Product Quantity9300 units

Reason for Recall

Mislabeled-9F labeled units actually contain 10F devices

Firm Action

The recalling firm telephoned their consignees on 5/10/06 to inform them of the problem and the need to recall their kits

Root Cause

Other

Distribution Pattern

The product was shipped to four consignees in FL, NY, CT, and Germany that repack the product into their own kits.