One Touch SureSoft Lancing Device in carton. Part Number 021-140-01, aka OT SureSoft Hospital, Gentle 200
FDA Recall Z-1165-2007 · Posted 2007-08-16
Advertisement
Recalling FirmLifescan Inc
Recall StatusTerminated
Product CodeFMK
Date Initiated2007-03-30
Date Terminated2007-12-03
Product Quantity632 cartons (200 units per carton)
Reason for Recall
Lancet tip may not fully retract and cause needle stick and exposure to used lancets.
Firm Action
The firm has conducted a health hazard evaluation, and has initiated a corrective action plan. The firm initated recall on March 30, 2007, and notifications were sent to hospital end users known to Lifescan and its direct accounts. In addition, a customer service script has been developed to address any questions from incoming users.
The firm has requested direct accounts to perform subrecalls on Lifescan's behalf to notify customers of the recall as needed.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Product was distributed nationwide to distributors and wholesale dealers.