Pressure Injectable Multi- Lumen CVC Kit

FDA Recall Z-1165-2018 · Posted 2018-03-05

Recall StatusTerminated
Product CodeFOZ
Date Initiated2017-11-14
Date Terminated2019-07-15
Product Quantity18,124 total

Reason for Recall

Products do not contain the Central Venous Catheter Placement Procedures Note or the Central Line Insertion Checklist, which is a tool intended to guide clinicians in the use of best practices during central line insertions.

Firm Action

On November 16, 2017, an Urgent Medical Device Notification titled "ARROW PSI and Pressure Injectable Multi-Lumen CVC Kits" was issued to customers informing them of the issue, and urging all to take the following actions: 1. Place a copy of this notification and the enclosed Central Line Insertion Checklist with each unit of affected product currently in your inventory. 2. Complete the enclosed Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or e-mail to [email protected]. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.

Root Cause

Labeling mix-ups

Distribution Pattern

Nationwide Distribution