Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

AcuMatch XLE Item Number / Product Description: 140-28-27 AM EHXL 0 LNR G 28MM 140-32-27 AM EHXL 0 LNR G 32MM 140-32-28 AM EHXL 0 LNR H 32MM 140-28-29 AM EHXL 0 LNR J 28MM 140-32-29 AM EHXL 0 LNR J 32MM 140-36-29 AM EHXL 0 LNR J 36MM 140-32-30 AM EHXL 0 LNR K 32MM 140-36-30 AM EHXL 0 LNR K 36MM 140-28-25 AM EHXL 0LNR E 28MM 140-28-26 AM EHXL 0LNR F 28MM 140-28-28 AM EHXL 0LNR H 28MM 140-36-28 AM EHXL 0LNR H 36MM 140-28-30 AM EHXL 0LNR K 28MM 142-28-24 AM EHXL 15 LNR D 28MM 142-28-26 AM EHXL 15 LNR F 28MM 142-32-27 AM EHXL 15 LNR G 32MM 142-32-28 AM EHXL 15 LNR H 32MM 142-36-28 AM EHXL 15 LNR H 36MM 142-36-29 AM EHXL 15 LNR J 26MM 142-28-29 AM EHXL 15 LNR J 28MM 142-32-29 AM EHXL 15 LNR J 32MM 142-28-30 AM EHXL 15 LNR K 28MM 142-32-30 AM EHXL 15 LNR K 32MM 142-36-30 AM EHXL 15 LNR K 36MM 142-28-25 AM EHXL 15LNR E 28MM 142-28-27 AM EHXL 15LNR G 28MM 142-28-28 AM EHXL 15LNR H 28MM 144-28-28 AM EHXL EXT LNR H 28MM 144-28-29 AM EHXL EXT LNR J 28MM 144-28-25 AM EHXL EXTLNR E 28MM 144-28-26 AM EHXL EXTLNR F 28MM 144-28-27 AM EHXL EXTLNR G 28MM 144-28-30 AM EHXL EXTLNR K 28MM 148-28-26 AM EHXL LAT LNR F 28MM 148-28-27 AM EHXL LAT LNR G 28MM 148-36-28 AM EHXL LAT LNR H 36MM 148-36-29 AM EHXL LAT LNR J 36MM 148-28-30 AM EHXL LAT LNR K 28MM 148-36-30 AM EHXL LAT LNR K 36MM 148-28-25 AM EHXL LATLNR E 28MM 148-28-28 AM EHXL LATLNR H 28MM 148-28-29 AM EHXL LATLNR J 28MM All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment.

FDA Recall Z-1165-2025 · Posted 2025-02-12

Recalling FirmExactech, Inc.
Recall StatusOpen, Classified
Product CodeOQI
Date Initiated2024-12-31
Product Quantity42

Reason for Recall

Due to the lack of the Ethylene Vinyl Alcohol (EVOH) barrier layer.

Firm Action

On 12/31/2024, the firm sent an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" letter via FedEx to customers initially informing them that the firm was recalling specific hip polyethylene liner lots that were inadvertently packaged without the ethylene vinyl alcohol (EVOH) layer in the innermost bag. An updated customer notification letter was communicated to affected customer via email on 01/17/2025. Customers are instructed to: -Immediate Discontinuation: Immediately discontinue use of all affected products -Quarantine and Return: Segregate and quarantine affected products and arrange for their return to Exactech. The Recall Confirmation Form section of this communication provides product return instructions. Clinical Guidance: 1. Patient Monitoring: Surgeons should regularly monitor patients with affected devices for potential risks such as device wear, failure, component cracking or fracture, new or worsening pain, bone loss and/or swelling, as per the instructions for use. 2. Diagnostic Considerations: Diagnostic imaging, such as X-rays, may be considered if device failure is suspected. 3. Revision Considerations: Revision of well-functioning devices is not recommended for patients. 4. Resources: Healthcare providers and patients with questions can access educational resources on our website exac.com/recall-information and use the device serial look-up tool. Public Contact - 352-377-1140 or email [email protected]

Root Cause

Nonconforming Material/Component

Distribution Pattern

U.S.: AR, CA, CO, FL, GA, HI, IL, IN, KS, KY, LA, MD, MN, NV, NY, OH, OK, PR,SC, TN, TX and VA O.U.S.: Argentina, Colombia, Germany, Guatemala, Japan and Spain

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑