Terminated

Zoll E Series Defibrillator with 12 Lead ECG, Automated external defibrillator, Software Version 3.00 and higher; Zoll Medical Corporation, Chelmsford, MA 01824.

FDA Recall Z-1167-2008 · Posted 2008-06-28

Recall StatusTerminated
Product CodeMKJ
Date Initiated2008-01-15
Date Terminated2017-02-06
Product Quantity5,018 units

Reason for Recall

Incorrect Patient Records: Patient records that have been stored, then later transmitted or printed may not be the correct record. Records that are transmitted or printed immediately following acquisition are correct.

Firm Action

Consignees were notified by an Urgent Device Correction letter sent on 1/14/08. The letter instructed users to complete and return a response form. Upon receipt of the response form, the recalling firm will have sent a software update. The letter also provided instructions for users to follow in order to avoid the software error before the software update was received. For additional information, contact 1-800-348-9011.

Root Cause

Software design