Fixone Biocomposite Anchor
FDA Recall Z-1167-2026 · Posted 2026-01-26
Advertisement
Recalling FirmAju Pharm Co., Ltd.
Recall StatusOpen, Classified
Product CodeMAI
Date Initiated2025-12-17
Product Quantity1,088 units
Reason for Recall
Reports of device failure, including anchor breakage, anchor bending, anchor pull out, suture breakage, and secondary packaging breakage.
Firm Action
An URGENT MEDICAL DEVICE RECALL notice dated 12/17/25 was mailed to consignees. The notification instructs consignees to discontinue use of affected devices, segregate affected devices, and return the product to Aju Pharm Medical or the authorized distributor. If product was further distributed the recall notification is to be forwarded for their awareness. Aju Pharm states that no additional action is required for patients who have already undergone surgery using affected devices. Consignees with any questions should contact +82-02-2630-0641 by phone or [email protected].
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide distribution in the states of CA & PR.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K233601Fixone Meniscal RepairAju Pharm Co., Ltd. · 2024-08-02K230892Fixone Biocomposite AnchorAju Pharm Co., Ltd. · 2023-06-09K222423Fixone All Suture AnchorAju Pharm Co., Ltd. · 2023-03-23K203523Fixone hybrid knotless anchorAju Pharm Co., Ltd. · 2021-12-25K193497Fixone Biocomposite Interference ScrewAju Pharm Co., Ltd. · 2021-01-05
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.