Aquilion 32 CT System; Model number TSX-101 A/K The product is an x-ray device that uses digital geometry to generate a three-dimensional image of the inside of an object from a large series of two-dimensional X-ray images take around a single axis of rotation
FDA Recall Z-1168-2010 · Posted 2010-03-19
Advertisement
Recalling FirmToshiba American Medical Systems Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2009-01-08
Date Terminated2012-06-01
Product Quantity9
Reason for Recall
Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large volumes of data are being reconstructed on the secondary console.
Firm Action
Customer notification was sent and a software upgrade will be issued to all customers affected by the field correction.
If you have questions, please call (800) 521-1968.
Root Cause
Component design/selection
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.