Terminated

Trident Hemispherical Acetabular Shells HA Solid BK 42mm- 74mm; Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size A insert; hip prosthesis component, Stryker Orthopaedics, Howmedica Osteonics Corp, Mahwah, NJ 07430

FDA Recall Z-1169-2008 · Posted 2008-06-12

Recall StatusTerminated
Product CodeMEH
Date Initiated2008-01-21
Date Terminated2012-09-25
Product Quantity129,312 total shells, all Trident hemispherical types

Reason for Recall

Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.

Firm Action

Recall notification letters were sent to Stryker Branches/agencies, OR Supervisors and Chief of Orthopaedics on 1/21/08. A Patient information sheet was sent on 2/4/08 to surgeons and hospitals. Per call with Center 2/4/08, recall letter to be revised and RES updated upon receipt. The letter was revised and sent on February 28, 2008 to include Trident PSL Acetabular shells. This letter was sent to Risk Management at hospitals and included the scope of the recall , the potential hazard and recommendation to physicians to monitor patients consistent with care for those receiving total hip replacement. It was also sent to surgeons and a revised letter to Stryker branches. A product acknowledgement form was included in all letters to indicate receipt of letter and quantity of product on hand, if applicable.

Root Cause

Process design

Distribution Pattern

Nationwide distribution.