Captivator II Single-Use Polypectomy Snares Outer box UPN, Inner Pouch UPN: M00561191 (Box 10), M00561190; M00561221 (Box 10), M00561220; M00561222 (Box 20), M00561220; M00561223 (Box 40), M00561220; M00561233 (Box 40), M00561230.
FDA Recall Z-1169-2021 · Posted None
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusOpen, Classified
Product CodeFDI
Date Initiated2021-01-19
Product Quantity335,788 units total (Domestic: 222,882, Foreign: 112,906)
Reason for Recall
Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
Firm Action
On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.
Root Cause
Process change control
Distribution Pattern
US nationwide distribution.