Terminated

62 RNR Series 2.5 microliter syringe without needle, Part#: 7632-01

FDA Recall Z-1170-06 · Posted 2006-06-22

Recalling FirmHamilton Co
Recall StatusTerminated
Product CodeJQW
Date Initiated2006-03-28
Date Terminated2006-12-21
Product Quantity36 units

Reason for Recall

There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe

Firm Action

On 3/28/06, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.

Root Cause

Other

Distribution Pattern

Worldwide-The product was distributed to 28 U.S. consignees and 3 foreign consignees. (Canada, Switzerland, Japan)