Sekure Chemistry Sekisui Diagnostics Acetaminophen L3K Assay, In-vitro Diagnostic Catalog Number: 506-10: R1: 1 x 10ml, R2: 2 x 10ml Catalog Number: 506-30: R1: 3 x 10ml, R2: 6 x 10ml In-Vitro quantitative measurement of Acetaminophen in serum and plasma used m the diagnosis and treatment of acetaminophen overdose toxicity.

FDA Recall Z-1171-2016 · Posted None

Recall StatusTerminated
Product CodeLDP
Date Initiated2016-01-22
Date Terminated2016-08-09
Product Quantity14,407 kits.

Reason for Recall

Crystal formation in the Color Reagent (R2) when left on board automated analyzers causing samples to under recover

Firm Action

Sekisui sent an Important Field Correction - Customer Notification letter dated January 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Discard all reagent in house with a current expiration date of 2016-09. 2. Update the expiration on all remaining kits on hand with the new expiration dates provided 3. When adding additional reagent to the analyzer, a new reagent wedge should be used. 4. If crystals are observed within reagent stored on-board, reagent should be discarded and replaced with fresh reagent. 5. Complete, sign and return the attached CONFIRMATION OF NOTIFICATION form by FAX to the number provided. This will indicate receipt of this Field Correction, act as confirm action of compliance with the actions,and allow for issuance of replacement material for expired product on hand (lots 47677, 47694, 47695). Questions contact Technical Services at 1-800-565-0265 or via email at [email protected]

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Distribution to the states of : AZ, CA, FL, IA, ID, IL, MA, , MN, MT and ND., and to the countries of Canada, France and Hong Kong.