Open, Classified

Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.

FDA Recall Z-1171-2024 · Posted 2024-02-21

Recall StatusOpen, Classified
Product CodeLLZ
Date Initiated2024-01-03
Product Quantity13

Reason for Recall

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Firm Action

On January 3, 2024, the firm notified customers via URGENT MEDICAL DEVICE CORRECTION letters. Fujifilm will offer to correct the formula that calculates the LV Mass (2D Bullet) in the Synapse database. Until the update is scheduled and completed, customers may continue to safely use the Pediatric Workflow modules listed in the notice as long as the LV Mass (2D Bullet) equation is not the main factor for a diagnosis. Customers should coordinate with the Fujifilm service engineer to schedule the correction of their device when contacted. Please contact Fujifilm's field action coordinator, Kotei Aoki, via phone at 765-246-2931 or email at [email protected], if you have any further questions regarding this field action.

Root Cause

Software design

Distribution Pattern

US domestic distribution nationwide. International distribution to Israel.