Terminated

Voluson 730 Expert, Voluson 730 Pro, Voluson 730 ProV with software versions 4.0.0, 4.0.1, 4.0.2, 4.0.3, 4.0.4, 4.0.5, 5.00, 5.0.1. The device is a general-purpose diagnostic ultrasound system.

FDA Recall Z-1172-06 · Posted 2006-06-27

Recall StatusTerminated
Product CodeIYN
Date Initiated2006-02-24
Date Terminated2007-01-06
Product Quantity3,936 (969 US and 2,967 OUS)

Reason for Recall

In the cardiac measurement section of the device the calculation of the 'PG Mean' (mean pressure gradient) leads to an erroneous result. It is recommended that customers with affected units do not use the measurement in question 'PG Mean' for any diagnostic or therapeutic decision, and 'PG Mean' values should not be included in any patient documentation.

Firm Action

An Urgent Medical Device Correction letter to customers 04/17/2006, states the product affected, provides a recommendation for affected units continued use until a software update is provided to the facility by a GE Healthcare Representative. Customers are recommended to not use the measurement in question 'PG Mean' for any diagnostic or therapeutic decision, and 'PG mean' values should not be included in any patient documentation.

Root Cause

Other

Distribution Pattern

Nationwide, Puerto Rico and OUS to include: Algeria, Argentina, Austria, Belgium, Bosnia and Herzegovina, Brazil, Republic of Bulgaria, Byelorussian SSR, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, DVU Australia, Ecuador, Egypt, Finland, France, Federal Republic of Germany, Greece, Hong Kong, Republic of Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Kuwait, Republic of Lithuania, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Kingdom of Saudi Arabia, Singapore, Republic of Slovenia, Spain, Sudan, Sweden, Switzerland, Taiwan, United Republic of Tanzania, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Venezuela, Vietnam, Yemen, Yugoslavia