OPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and Lite

FDA Recall Z-1173-2024 · Posted 2024-02-22

Recall StatusOpen, Classified
Product CodeQFV
Date Initiated2024-01-05
Product Quantity1,469 total units

Reason for Recall

OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.

Firm Action

On January 2, 2024, the firm sent letters to field representatives to be given to affected physicians. Summary of the safety communication can be found online: https://impulse-dynamics.com/safety-information/ The firm recommends all patients with implant card showing model CCM X11, who have experienced the "A9" charge error, charge their implanted device to no more than 75% of full capacity to circumvent the issue. See link above for instructions. If a patient encounters an A9 error code, CCM therapy has been suspended and must be manually reset. Patients should schedule time with an Impulse Dynamics representative or their physician to manually reset their device. The firm is working on a software update to address this issue. Patients and clinicians may determine if a specific device is affected by providing the serial number of any specific device to the Technical Support Hotline at 866-312-5370.

Root Cause

Software design

Distribution Pattern

Domestic US distribution nationwide. International distribution to Austria Croatia Czech Republic Germany Italy Macau Poland Romania Russia Spain Sweden.