Terminated

Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)

FDA Recall Z-1175-2024 · Posted 2024-02-22

Recalling FirmAbiomed, Inc.
Recall StatusTerminated
Product CodeOZD
Date Initiated2021-03-01
Date Terminated2025-03-06
Product Quantity4,463 (US)

Reason for Recall

Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.

Firm Action

This is a retrospective report for the March 2021 Technical Bulletin, which described a Product Update; specifically the AIC Version 8.5 Software Update. The software update began rollout on March 4, 2021. According to the firm, Version 8.5 of the software contained bug fixes for the issues seen in version 8.4. The technical bulletin was posted to HeartRecovery.com on March 5, 2021.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution.

Official Source

View FDA Recall →