Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)
FDA Recall Z-1175-2024 · Posted 2024-02-22
Advertisement
Recalling FirmAbiomed, Inc.
Recall StatusTerminated
Product CodeOZD
Date Initiated2021-03-01
Date Terminated2025-03-06
Product Quantity4,463 (US)
Reason for Recall
Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
Firm Action
This is a retrospective report for the March 2021 Technical Bulletin, which described a Product Update; specifically the AIC Version 8.5 Software Update. The software update began rollout on March 4, 2021. According to the firm, Version 8.5 of the software contained bug fixes for the issues seen in version 8.4. The technical bulletin was posted to HeartRecovery.com on March 5, 2021.
Root Cause
Software design
Distribution Pattern
US Nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.