Terminated

Roche Calcium Reagent Kit for use on the Cobas MIRA Chemistry System; in vitro diagnostic; Cat. No. 04585984160.

FDA Recall Z-1176-06 · Posted 2006-06-27

Recall StatusTerminated
Product CodeCIC
Date Initiated2006-06-07
Date Terminated2008-04-23
Product Quantity933 kits

Reason for Recall

Falsely low patient and control results may be reported when the reagent has been on-board for less than the 8 hour use period specified in the labeling.

Firm Action

Consignees were notified via letter dated 6/6/06 and sent via UPS to discontinue use of these lots and to destroy them. Distributors are being asked to notify their customers.

Root Cause

Other

Distribution Pattern

Nationwide.