PhaSeal Injector Luer Lock, Item No. 31, US Order No. 10310

FDA Recall Z-1179-2007 · Posted 2007-08-22

Recalling FirmCarmel Pharma, Inc.
Recall StatusTerminated
Product CodeFPA
Date Initiated2007-06-26
Date Terminated2013-03-05
Product Quantity6.1 million for Z-1178-1181-2007 recalls

Reason for Recall

Risk of Leakage: When the PhaSeal Protector and Injector are used with drug vials with thick rubber stoppers which have a concave upper surface and a flat bottom surface there is a potential for leakage. These rubber stoppers are made of a harder rubber formulation and may be more difficult to penetrate.

Firm Action

On or about 6/15/2007, the firm sent advisory notice letters to the end users and the instructions for use have been updated by adding a precaution statement regarding the use of thick rubber stoppers. The firm is also updating the customer training program to include how to handle drug vials with thick concave rubber stoppers.

Root Cause

Other

Distribution Pattern

Product was distributed throughout the US and Canada