da Vinci S Surgical System, Model number IS2000 Revision number A51 P5; endoscopic surgical control system, Intuitive Surgical, Inc., Sunnyvale, CA 94086
FDA Recall Z-1180-2008 · Posted 2008-06-12
Advertisement
Recalling FirmIntuitive Surgical, Inc.
Recall StatusTerminated
Product CodeNAY
Date Initiated2007-12-27
Date Terminated2008-08-15
Product Quantity9 units
Reason for Recall
Delay in responding: In certain circumstances, the device may not respond immediately to a user's command such as master clutch or camera control.
Firm Action
Beginning 12/28/2007, Intuitive Surgical Field Service Representatives visited each site to correct the issues. A notification letter was provided to all sites.
Root Cause
Software design
Distribution Pattern
USA Distribution: TX, NJ, WA, NV, HI, MI, NH and CA
Related FDA 510(k) Records