da Vinci S Surgical System, Model number IS2000 Revision number A51 P5; endoscopic surgical control system, Intuitive Surgical, Inc., Sunnyvale, CA 94086

FDA Recall Z-1180-2008 · Posted 2008-06-12

Recall StatusTerminated
Product CodeNAY
Date Initiated2007-12-27
Date Terminated2008-08-15
Product Quantity9 units

Reason for Recall

Delay in responding: In certain circumstances, the device may not respond immediately to a user's command such as master clutch or camera control.

Firm Action

Beginning 12/28/2007, Intuitive Surgical Field Service Representatives visited each site to correct the issues. A notification letter was provided to all sites.

Root Cause

Software design

Distribution Pattern

USA Distribution: TX, NJ, WA, NV, HI, MI, NH and CA