Quick Connects. Steris Corporation. Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT. Each Quick Connect contains a flow unit consisting of tubing, tethers, and adapters. For sterilant and rinse water used in endoscope washers.

FDA Recall Z-1180-2016 · Posted None

Recalling FirmSteris Corporation
Recall StatusTerminated
Product CodeMED
Date Initiated2015-12-04
Date Terminated2016-08-26
Product Quantity809 units

Reason for Recall

The firm discovered that one of the hysteroscope models listed for processing in four existing Quick Connects has one inlet port rather than two inlet ports as indicated in the Quick Connect processing instructions.

Firm Action

On 1/18/2016 the firm sent Voluntary Field Correction NOtices to their customers.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution. US nationwide, Australia, Bahrain, Bulgaria, Canada, Greece, Italy, Malaysia, Mexico, Qatar, Russia, Spain, and United Kingdom.