Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.
FDA Recall Z-1183-2018 · Posted None
Advertisement
Recalling FirmQuidel Corporation
Recall StatusTerminated
Product CodePSZ
Date Initiated2017-12-15
Date Terminated2019-10-21
Product Quantity482 units
Reason for Recall
The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.
Firm Action
Quidel sent and Urgent Product Recall letter dated December 15, 2017.
The letter instructs the user to discontinue use of the device until new firmware can be installed. For further questions, please call (800) 874-1517.
Root Cause
Software Manufacturing/Software Deployment
Distribution Pattern
US Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.