Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733), for in vitro diagnostic use.
FDA Recall Z-1184-2008 · Posted 2008-08-20
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeJJE
Date Initiated2007-10-08
Date Terminated2012-01-11
Product Quantity111 units in the US; 9 units in Canada
Reason for Recall
Incorrect results: There is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027)
and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range.
Firm Action
Beckman Coulter mailed a Product Corrective Action (PCA) letter the week of
October 8, 2007 to the customers who have the affected UniCel Dxl 800
Access lmmunoassay systems advising that there is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027)
and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range. The letter further advises that Beckman Coulter will contact the customer to arrange a date to remove the modificatlons but until the modification is completed, the system should be operated in a
room temperature of 20C to 30C. This can be acomplished by monitoring
and recording room tamperature in the area of the instrument. Customers are also asked to review historical quality control records to ensure assay performance is within their laboratory specification and that patient results should be reviewed in context of other diagnostics tests and the clinical presentations of the patient. A response form is included.
Root Cause
Device Design
Distribution Pattern
Nationwide and Canada