Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of anti-thyroglobulin antibodies (TgAb) in human serum on specific TOSOH AIA System analyzers. Measurements obtained by this device aid in the diagnosis of autoimmune thyroid disorders, such as Graves disease and Hashimotos disease. Comment to IU: Please note that the Calibrator set is not part of the IU but required for the assay.
FDA Recall Z-1187-2021 · Posted None
Advertisement
Recalling FirmTosoh Bioscience Inc
Recall StatusTerminated
Product CodeJZO
Date Initiated2012-02-23
Date Terminated2021-07-08
Product Quantity34 boxess/2 sets/6 bottles
Reason for Recall
Some combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable.
Firm Action
On 7/24/2012 the firm sent recall notificaiton letters to their customers.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide distribution in the states of CA, FL, GA, NY, OH & OR.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.