Left Long Tactile Probe, Part #962009S, Medtronic Navigation, Louisville, CO 80027. Trackable hand held instruments used in spine surgeries for preparation of the pedicle. These instruments are used in conjunction with the Medtronic Navigation StealthStation platform and various software applications.
FDA Recall Z-1189-2012 · Posted 2012-03-08
Advertisement
Recalling FirmMedtronic Navigation, Inc.
Recall StatusTerminated
Product CodeHAW
Date Initiated2012-02-06
Date Terminated2012-03-26
Product Quantity6 units
Reason for Recall
When subjected to bending, twisting, and malleting, the probe tips break off when used in the sacrum for which they are not indicated for use. These probes are approved for use in the lumbar vertebra.
Firm Action
An Urgent - Product Safety Notice customer letter was sent via FED EX on 2/6/12 and 2/6/12 to direct consignees in the US identifying the affected product and the issue with it. The letter states that Medtronic Navigation is removing the product from clinical use and will provide a sales credit to customers. In the letters to Medtronic field personnel, it states that local sales and service personnel will be identified to remove the affected product from customer sites and return them for scrapping. Customers are to contact a sales representative if they have any questions regarding the letter.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution -- US and Republic of South Korea.