BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (4.5 ml x 13 x 75 mm) Catalog Number: 369714 Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations

FDA Recall Z-1189-2017 · Posted None

Recall StatusTerminated
Product CodeJKA
Date Initiated2017-01-16
Date Terminated2018-02-07
Product Quantity1,488,800 tubes

Reason for Recall

A portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount of buffered sodium citrate additive, an anticoagulant.

Firm Action

BD sent an Urgent Product Recall letter dated January 16, 2017 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the Customer Recall Response Form and fax the completed form back to BD at 1-844-388-5534 or email the completed form to [email protected]. For questions call 1-844-265-7408.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - US Nationwide and the countries of Canada, Mexico, Indonesia

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.